LC4
25
For each parameter set, the control parameters Xp, Tn and Tv
may be individually modified and stored (see page 19).
Preference:
Quick access to parameters determined during previous
applications.
Factory-preset parameter sets for the following
controlled processes:
Set 1: Xp = 4 °C
Tn = 150 s
Tv = 10 s
Directly heated liquid bath
Filling volume: 4.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 2000 W
Set 2 Xp = 10 °C
Tn = 300 s
Tv = 25 s
Directly heated liquid bath
Filling volume: 4 litres of Thermal H
Setpoint temperature: 150 °C
Heater capacity: 2000 W
Set 3 Xp = 20 °C
Tn = 1800 s
Tv = 165 s
Heating hood model JH2000 with 2000 ml round bottom
flask
Filling volume: 2 litres of water
Setpoint temperature: 70 °C
Heater capacity: 480 W (stage III)
Set 4 Xp = 35 °C
Tn = 1800 s
Tv = 130 s
Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 150 W (stage II)
Set 5 Xp = 15 °C
Tn = 1550 s
Tv = 120 s
Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of Thermal H
Setpoint temperature: 100 °C
Heater capacity: 150 W (stage II)